Question 2 of 7: X-rays, Attenuation and Absorption in Matter, Dosimetry
Nivaar worked solution (AI-drafted; not reviewed by a licensed engineer)
Notes on this paper
Paper format. 98-Phys-B1 Radiation Physics, National Examination
December 2013 — a three-hour open-book examination in which any
non-communicating calculator is permitted. The cover page states that all seven
questions must be attempted (no choose-N-of-M here) for a total of 100 points, and
invites the candidate to submit a written statement of any assumptions made where a
question is open to interpretation. This licence is used below in Question 2(c) (the
photon-production law assumed for the current change) and Question 5(a) (the numeric value
behind the news item's "five times the acceptable exposure" claim).
Reference texts. K. S. Krane, Introductory Nuclear Physics
(nuclear masses and binding energy, radioactive decay, fission); F. H. Attix,
Introduction to Radiological Physics and Radiation Dosimetry (X-ray production and
bremsstrahlung spectra, photon interactions — photoelectric effect, Compton
scattering, pair production, gamma detectors); J. R. Cember and T. E. Johnson,
Introduction to Health Physics, 5th ed. (dose equivalent, internal dosimetry and
effective half-life, shielding, fission-product hazards); J. E. Turner, Atoms,
Radiation, and Radiation Protection, 3rd ed. (radiation interactions with matter,
health-physics standards).
Question 2: X-rays, Attenuation and Absorption in Matter, Dosimetry (20 marks)
Given. Tube voltage 60 kV rising to 75 kV; path length 20 mm muscle +
3.5 mm fat + 6.5 mm bone; exposure time fixed at 0.5 s in both cases; mean photon energy
assumed to be one-third of the peak (accelerating) energy; the attenuation-coefficient table
above, which conveniently tabulates exactly the two required mean energies (60/3 = 20 keV
and 75/3 = 25 keV).
Find. (a) the maximum photon energy; (c) the ratio of applied currents
that keeps the film exposure (photon count) unchanged; (d) whether the resulting patient
dose rises, falls, or stays the same; (e)–(g) short conceptual answers on dosimetry
and attenuation.
Approach. Bremsstrahlung's maximum photon energy equals the electron's
kinetic energy at the target, $E_{\max}=eV$, giving (a) directly. For (c)–(d), take
the standard radiographic simplification that photon quantity (the number of
photons leaving the target) is controlled by mAs (current × time) essentially
independent of kVp, while photon quality (mean energy and hence penetration) is
set by kVp — this is the textbook "mAs controls quantity, kVp controls quality"
convention, flagged explicitly since the exam's own instructions invite a stated assumption.
The exponential attenuation law and the given coefficient table then convert between photon
counts at the film for the two mean energies.
Part (a) — maximum photon energy. Bremsstrahlung photons cannot
carry more energy than the accelerated electron that produces them, so the spectrum's upper
edge is set by the tube voltage:
$$\boxed{E_{\max} = eV = 60\text{ keV}}$$
Part (b) — spectrum sketch. The bremsstrahlung spectrum, as
filtered by the tube window, is continuous, rising from zero intensity at low photon energy
(the softest photons are absorbed in the window), peaking at roughly
one-third to one-half of $E_{\max}$, and falling to zero intensity at $E_{\max}=60$ keV (no
photon can exceed the electron's kinetic energy); with no target material specified, no
discrete characteristic lines are superimposed.
Figure 1 — bremsstrahlung spectrum (as emerging through the tube window): continuous, peaking near 20–25 keV and falling to zero intensity at the 60 keV cut-off.
Part (c) — current change to hold film exposure constant. Under
the "mAs sets quantity" assumption, the photon count produced in time $t$ is $N_0 \propto
i\,t$, independent of $V$; the count reaching the film after traversing the tissue stack is
$N_{\text{film}} = N_0\,T(E)$ with transmission $T(E)=e^{-\sum \mu_j(E)x_j}$. Summing the
table's coefficients over the given thicknesses ($x_{\text{muscle}}=2.0$ cm,
$x_{\text{fat}}=0.35$ cm, $x_{\text{bone}}=0.65$ cm) at the two mean energies:
$$\textstyle\sum \mu x\,(20\text{ keV}) = 0.808(2.0)+0.488(0.35)+4.542(0.65) = 4.739$$
$$\textstyle\sum \mu x\,(25\text{ keV}) = 0.594(2.0)+0.379(0.35)+3.058(0.65) = 3.308$$
$$T(20) = e^{-4.739} = 8.75\times10^{-3}, \qquad T(25) = e^{-3.308} = 3.66\times10^{-2}$$
Equating film counts $i_1 t\,T(20) = i_2 t\,T(25)$ (exposure time cancels since it is the
same 0.5 s in both cases):
$$\boxed{\frac{i_2}{i_1} = \frac{T(20)}{T(25)} = \frac{8.75\times10^{-3}}{3.66\times10^{-2}}
\approx 0.239}$$
The applied current must be reduced to about 24% of its original value
(roughly a 76% cut) — the 75 kV beam is so much more penetrating through this tissue
stack that far fewer photons need to be produced to deliver the same count to the film.
Part (d) — effect on patient dose. Dose is proportional to the
number of photons absorbed by the patient (not transmitted) times their energy.
The photon count entering the patient scales with $i\,t$ (same law as part (c)), so with
$i_2/i_1 \approx 0.239$ from above and the absorbed fraction $1-T(E)$:
$$\frac{\text{Dose}_2}{\text{Dose}_1} =
\frac{(i_2 t)\,[1-T(25)]\,E_2}{(i_1 t)\,[1-T(20)]\,E_1}
= \frac{0.239\times(1-0.0366)\times25}{1\times(1-0.00875)\times20}$$
$$\boxed{\frac{\text{Dose}_2}{\text{Dose}_1} \approx 0.29}$$
The dose at 75 kV (with the current reduced to keep the film count fixed) is
lower — about 29% of the original 60 kV dose. Even though each
photon of the 75 kV beam (25 keV mean) carries 25% more energy than one of the 60 kV beam
(20 keV mean), roughly four times fewer photons now enter the
patient, and that reduction dominates.
Part (e) — measuring absorbed dose in practice. Directly, with a
calibrated dosimeter placed at the point of interest: an ionization chamber or a
thermoluminescent dosimeter (TLD) reads exposure/air kerma, which is then converted to
tissue absorbed dose via known conversion (f-)factors; in diagnostic radiography this is
usually reported as entrance skin dose or dose-area product from a calibrated chamber built
into the collimator, cross-checked against the known kVp/mAs/distance technique factors.
Part (f) — dominant interaction mode. At 20–30 keV, the
photoelectric effect dominates. The table itself is the evidence: bone's
coefficient (Z of Ca, P far exceeds soft tissue) is about 4–6× higher than
muscle's at the same energy, and every coefficient falls steeply with increasing energy — both
signatures of a cross-section scaling roughly as $Z^{3\text{-}4}/E^3$, which is the
photoelectric effect; Compton scattering's cross-section is nearly energy- and
Z-independent per electron and would not produce either trend.
Part (g) — why bone attenuates more. Bone's mineral content (Ca,
P) gives it a much higher effective atomic number than the light elements (H, C, O, N)
dominating muscle and fat, and the photoelectric cross-section — dominant at these
energies per part (f) — scales steeply with atomic number ($\propto Z^{3\text{-}4}$).
Bone's higher physical density also raises its electron density per unit volume. Both
effects compound, which is exactly why bone appears bright on a radiograph.
Check: Part (c)/(d) assumes the standard radiographic simplification
that photon quantity tracks mAs (current × time) essentially independently
of kVp, per the exam's invitation to state assumptions; a more detailed bremsstrahlung yield
model (output scaling roughly with $i\,t\,V^2$) would shift the numeric current ratio but
not the qualitative conclusion that raising kVp lets mAs (and hence dose) drop sharply
— this is exactly the clinical "increase kVp, decrease mAs" technique used to reduce
patient dose while holding film density constant.
Question 2 — results
Quantity
Value
(a) Maximum photon energy
60 keV
(c) Required current ratio $i_2/i_1$ (75 kV vs 60 kV)